Publication Type : Journal Article
Publisher : International Journal of Advanced Research in Chemical Science (IJARCS)
Source : International Journal of Advanced Research in Chemical Science (IJARCS), vol. volume 3, no. Issue 2, 2016
Campus : Amritapuri
School : School of Arts and Sciences
Center : Research & Projects
Department : Chemistry
Verified : Yes
Year : 2016
Abstract : N 1 , N2 -bis(salicylidine)butane-1,4-diamine (SAB ) based PVC membrane sensor was fabricated for Nd(III). The developed sensor showed nernstian response in the concentration range of 1.0×10-6 to 1.0×10-1 M. Nernstian slope of the proposed sensor was found to be 20.6±0.3 mV/decade in the pH range 3.0 to 7.5. The developed sensor displays a response time of lt;15 s. The proposed sensor can be used over a period of 7 weeks without any significant changes in its Nerstian behaviour. The sensor showed good selectivity for neodymium when compared with other metal ions such as Na+ , K+ , Mg2+, Ca2+, Pb2+, Ag+ , Pr3+ , Sm3+, La3+, and Gd3+. This electrode was used as an indicator electrode in the potentiometric titration of neodymium ions with EDTA and was also applied in the determination of neodymium in binary mixture.
Cite this Research Publication : Dr. Beena S. and Krishnapillai Girish Kumar, “PVC Membrane Sensor for Nd (III) Based on N 1 , N 2 -Bis (Salicylidine) Butane-1, 4 -Diamine”, International Journal of Advanced Research in Chemical Science (IJARCS), vol. volume 3, no. Issue 2, 2016