Publication Type : Journal Article
Publisher : Indian Journal of Orthopedics
Source : Indian Journal of Orthopedics 2013;47(5):519-22
Url : https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3796928/
Campus : Faridabad
Year : 2013
Abstract : Management of recurrent dislocation of total hip arthroplasty is often a challenging and daunting task. Re-revision of such a total hip prosthesis may not be easy as the removal of a well-fixed, fully coated stem is extremely difficult. We managed to salvage instability in three revision hip cases in which the fully coated stem had subsided by using a bioball universal neck adapter without changing the femoral stem or acetabular cup.
Cite this Research Publication : Vaishya R,Sharma M. Bioball modular neck adapter as a salvage for failed revision total hip arthroplasty.Indian Journal of Orthopedics 2013;47(5):519-22