Publication Type : Conference Paper
Publisher : IOP Publishing
Source : IOP Conference Series: Materials Science and Engineering Volume 243 Issue 1 Pages 012010, 2017
Url : https://iopscience.iop.org/article/10.1088/1757-899X/243/1/012010/meta
Campus : Amritapuri
School : School of Engineering
Department : Mechanical Engineering
Year : 2017
Abstract : Wall shear stress is a crucial parameter used for blood damage analysis, and typically a value of 400 Pa is set as a limit. Tip clearance is a major factor contributing to hemolysis and pump efficiency. In this study, different tip gap configurations are used to analyse the wall shear stress developed on the blade surface of a constant thickness blade design, and a varying thickness blade design using CFD analysis. It was found that, for a particular geometry, as the clearance gap reduces, flow rate over the high wall shear stress area decreases even though the high wall shear stress span is found to extend. For each design, the optimum clearance gap is iteratively attained, keeping the maximum WSS as a limiting factor. Thus a better pump designs is obtained, whose leakage flow patterns are lower than that of the initial design, hence also leading to higher pump efficiency.
Cite this Research Publication : S Sarath, R Vikas, "Effect of tip clearance on wall shear stress of an axial LVAD", IOP Conference Series: Materials Science and Engineering Volume 243 Issue 1 Pages 012010, 2017