Publication Type : Journal Article
Publisher : Annals of Cardiac Anaesthesia
Source : Annals of Cardiac Anaesthesia 2008;11(2):90-95
Campus : Faridabad
Year : 2008
Abstract : The present study compares the cardiac output (CO) estimated by a new, non-invsive continuous Doppler device (Ultrasonic cardiac output monitor-USCOM) with that by bolus thermodilution technique (TD). Thirty post off-pump coronary artery bypass graft surgery patients were studied in this prospective non-randomozed study. Right heart CO estimation by USCOM and TD was performed and measured in quadruplet. A total of 120 paired observations were made. The mean CO was 4.63 and 4.76 L/min as estimated by TD and USCOM respectively. For TD and USCOM, the CO had a mean bias (tendency of one technique to differ from other) of -0.13 L/min and limits of agreement (mean bias-2SD) at -0.86 and 0.59 L/min. The study reveals very good agreement between the values of CO estimated by USCOM and TD.
Cite this Research Publication : Cardiac output estimation after OPCAB: A comparison of two different techniques. Annals of Cardiac Anaesthesia 2008;11(2):90-95